1. Home
  2. /
  3. X-ray components
  4. /
  5. X-ray tubes
  6. /
  7. Housing
  8. /
  9. Housing C40 for X-ray tube, IAE

Housing C40 for X-ray tube, IAE

Rotating anode X-ray tube C40 unit specifically designed for anode diameter inserts ranging from 70 mm to 80 mm, for medium/high duty radiographic procedures.

Housing C40 for X-ray tube, IAE

SPECIFICATIONS Housing C40
Overall length 490 mm
Maximum diameter 152 mm
Tube assembly net weight 16 kg
Nominal X-ray tube assembly voltage 150 kV
Maximum tube assembly heat content 900 kJ (1250 kHU)
Maximum continuous heat dissipation 180 W (14400 HU/min)
Minimum tube assembly inherent filtration 1.2 mm Al/ 75 kV
Maximum leakage radiation at 1 m from focal spots 0.44 mGy/h
    CHARACTERISTICS Housing C40

  • Lead lined aluminum body.
  • Two devices are present for thermal safety:
    1 a bimetallic thermal switch, internally assembled, series connected with stator common cable
    2 a pressure microswitch, installed on cathode end, activated by bellow expansion