Housing C100XT for X-ray tube, IAE

Special rotating anode X-ray tube C100XT unit, designed for digital and vascular applications. Outstanding thermal dissipation without heat exchanger.

Housing C100XT for X-ray tube, IAE

SPECIFICATIONS Housing C100XT
Overall length 524 mm
Maximum diameter 190 mm
Tube assembly net weight 29 kg
Nominal X-ray tube assembly voltage 150 kV
Maximum tube assembly heat content 1500 kJ (2000 kHU)
Maximum continuous heat dissipation 1000 W (80000 HU/min)
Minimum tube assembly inherent filtration 1.2 mm Al/ 75 kV
Maximum leakage radiation at 1 m from focal spots 0.44 mGy/h
    CHARACTERISTICS Housing C100XT

  • Lead lined aluminum body.
  • Three devices are present for thermal safety:
    1 a bimetallic thermal switch, fitted externally on the anode end
    2 a bimetallic thermal switch, internally assembled, series connected with stator common cable
    3 a pressure microswitch, installed on cathode end, activated by bellow expansion