Bucky Tables
- Attractive and elegant
- Photocell activation
- Smooth and silent
Advantages
- Maximum excursion available on all movements
- “Dead Man” double safety on all movements
- Light structure
- Easy and fast to install and maintain
- Structural resistance proved with severe tests
AIN Standard Bucky Table
Additional information | |
Weight | 152 kg |
Maximum patient weight | 200 kg |
Available image detectors: | |
Potter Bucky | |
Longitudinal movement | 550 mm |
Film-tabletop distance | 70 mm |
Cassette size | up to 43×43 cm |
Flat Panel | |
Longitudinal movement | depending on flat panel model |
Image plan-tabletop distance | depending on flat panel model |
See the scheme » AIN Standard Bucky Table »
UPPER – Elevating Bucky Table
Additional information | |
Weight | 220 kg |
Maximum patient weight | 200 kg |
Available image detectors: | |
Potter Bucky | |
Longitudinal movement | 550 mm |
Film-tabletop distance | 67 mm |
Cassette size | up to 43×43 cm |
Flat Panel | |
Longitudinal movement | depending on flat panel model |
Image plan-tabletop distance | depending on flat panel model |
See the scheme » UPPER – Elevating Bucky Table »
Accessories
Pressing Band
Band with suitable support to fit the aluminium profiles.
Allows to press soft tissue and facilitates some patient positioning operations.
Handgrips
Mobile supports to help tabletop adjustment.
Handles
Mobile and removable supports to help patient stability during examination.
Lateral cassette holder
Support with mechanical fittings to position cassettes perpendicularly to tabletop surface and allow lateral exposures.
A.E.C.
Ready for installation of A.E.C. chambers.
Electrical and Functional Features
Maximum patient weight | 200 Kg |
Up/down speed | 17 mm/sec (Upper only) |
Movement command | Ergonomic foot levers on front side |
Brake commands | Ergonomic foot levers on front side |
Power supply | 230 Vac (±10%) |
Frequency | 50/60 Hz |
Max adsorption | 0,9 kVA |
Installation type | Permanent |
Classification | Active Class IIb Medical Device |
Applied Part | Type B |
Working Environment | Hospital and Clinic |
Operative enviroment | Temperature range: 10°C to 50°C Relative Humidity: max 80% Atmospheric Pressure: 700 hPa to 1100 hPa |
Mechanical Structure
The mechanical structure of all elements is manufactured following a careful plan to grant quality, reliability, strength and resistance.
The installation is fast and easy, motorized movements have soft-start and soft-stop to allow a smooth adjustment. All safeties are “dead man”.
UPPER table has variable height for radiological diagnostic purposes.
The lower position of the table helps positioning of patients with reduced mobility and the wide movement of the tabletop allows different examination.
ARCOM Bucky tables are the ideal solution for common and uncommon X-ray investigations.
Features could change with no advice.