Housing C52 for X-ray tube, IAE
Rotating anode X-ray tube C52 unit specifically designed for anode diameter inserts ranging from 70 mm to 90 mm, for medium/high duty radiographic procedures.
SPECIFICATIONS Housing C52 | |
Overall length | 484 mm |
Maximum diameter | 170 mm |
Tube assembly net weight | 21 kg |
Nominal X-ray tube assembly voltage | 150 kV |
Maximum tube assembly heat content | 1280 kJ (1700 kHU) |
Maximum continuous heat dissipation without fan | 230 W (18400 HU/min) |
Maximum continuous heat dissipation with fan | 370 W (29600 HU/min) |
Minimum tube assembly inherent filtration | 1.2 mm Al/ 75 kV |
Maximum leakage radiation at 1 m from focal spots | 0.44 mGy/h |
- CHARACTERISTICS Housing C52
- Lead lined aluminum body.
- Three devices are present for thermal safety:
1 a bimetallic thermal switch, fitted externally on the anode end
2 a bimetallic thermal switch, internally assembled, series connected with stator common cable
3 a pressure microswitch, installed on cathode end, activated by bellow expansion