BReeZe EBP-D02 – Base Radiology System
- Analogic or digital detector
- Multiple detectors compatibility
- Ergonomic controls
- Motorized movements
Base Radiology System (BRS), available in various configurations with fixed or variable focal distance, manual or motorized positioning, ELM brakes. Designed for a practical use by any operator, in combination with an X-Ray transparent stretcher, allows a wide range of common examinations even in smallest cabinets.
A flexible system with several available uses: multiple positioning of a single patient achievable with easy adjustment and a few movements. Quick installation and easy to use, high compatibility with all types of X-Ray tubes, for maximum system flexibility. All BRS versions support stationary grid devices, Potter Bucky, and docking stations for several Flat Panels.
Variable Focal Distance
Technical Features
Motorized vertical movement | 1450 mm |
Minimum height from floor of X-Ray focus | 365 mm |
Swivel C-shaped arm | |
Rotation (motorized) | 30° left, 120° right |
Focal Distance | variable from 1000 to 1800 mm |
Angle indicator (inclinometer) | |
X-Ray tube-collimator support group | |
Rotation (motorized) | 180° left, 180° right with ELM Brake |
Angle indicator (inclinometer) | |
Manual collimator | |
Image detector support | |
Maximum image size | 43×43 cm |
X-Ray fine line grid | |
Accessories support | |
Rotation | 90° left, 90° right |
Angle indicator (inclinometer) |
See the scheme » Dimensions of BReeZe Variable Focal Distance »
Fixed Focal Distance
Technical Features
Motorized vertical movement | 1350 mm |
Minimum height from floor of X-Ray focus | 400 mm |
Angle indicator (inclinometer) | |
Swivel C-shaped arm | |
Rotation (manual) | 30° left, 120° right |
Focal Distance | 1400 mm |
Angle indicator (inclinometer) | |
X-Ray tube-collimator support group | |
Rotation (manual) | 180° left, 180° right |
Angle indicator (inclinometer) | |
Manual collimator | |
Image detector support | |
Maximum image size | 43×43 cm |
X-Ray fine line grid | |
Accessories support | |
Rotation | 45° left, 45° right |
Angle indicator (inclinometer) |
See the scheme » Dimensions of BReeZe Fixed Focal Distance »
Variable Focal Distance
Technical Features
Motorized vertical movement | 1350 mm |
Minimum height from floor of X-Ray focus | 400 mm |
Angle indicator (inclinometer) | |
Swivel arm | |
Rotation (motorized) | 30° left, 120° right |
Focal Distance | variable from 1000 to 2000 mm |
X-Ray tube-collimator support group | |
Rotation (motorized) | 180° left, 180° right with ELM Brake |
Angle indicator (inclinometer) | |
Manual collimator | |
Image detector support | |
Maximum image size | 43×43 cm |
X-Ray fine line grid | |
Accessories support | |
Rotation | 45° left, 45° right |
Angle indicator (inclinometer) |
See the scheme » Dimensions of BReeZe Variable Focal Distance »
System Components*
*Different devices are available under ARCOM approval
X-Ray Tubs
Wide choice of X-ray tubes to best fulfil any technical and diagnostic requirement.
We select the best manufacturers around the world to fit in every budget, being confident of quality and reliability. Heavy duty? Specialized operation? Emergency room?
We have the right solution for every request!
X-Ray Generator
An X-Ray generator is the heart of every radiological system.
Features and performances are important as an easy-to-use interface, involving fast and easy installation requirements.
Good generators are forms of art that require patience, experience and a deep knowledge of cutting edge technology in radiology field.
MODEL | Genesis 40 DT | Genesis 50 DT | Genesis 60 DT |
Power | 40kW | 50kW | 65kW |
RADIOGRAPHY MODE | |||
Nominal High Voltage | 150kV | 150kV | 150kV |
RX Tube Highest Current | 650 mA | 650 mA | 800 mA |
max mA @ max kV | 650 mA @ 75kV 400 mA @ 100kV 300 mA @ 125kV 250 mA @ 150kV |
650 mA @ 75kV 500 mA @ 100kV 400 mA @ 125kV 300 mA @ 150kV |
650 mA 800 mA @ 75kV 650 mA @100kV 500 mA @ 125kV 400 mA @ 150kV |
Power Supply | 400 V three-phase – 50Hz / 60Hz | ||
Inverter frequency | 120 kHz | ||
Automatic Exposure Control | Ramp function and pulse-signal supported | ||
APR Technique | Virtually unlimited | ||
80kW power available on request! |
This system is completed Dose Area Product Measuring System VacuDAP and Ionization Chambers AEC Sensor, the use of which is mandatory according to article 17 the Law of Ukraine «On Protection of People against lonizing Radiation».