Housing C31 for X-ray tube, IAE

Rotating anode X-ray tube C31 unit specifically designed for mobile equipment.

Suitable for max. 72 mm anode diameter inserts.

Housing C31 for X-ray tube, IAE

SPECIFICATIONS Housing C31
Overall length 400 mm
Maximum diameter 134 mm
Tube assembly net weight 15 kg
Nominal X-ray tube assembly voltage 150 kV
Maximum tube assembly heat content 500 kJ (670 kHU)
Maximum continuous heat dissipation 120 W (9600 HU/min)
Minimum tube assembly inherent filtration 1.2 mm Al/ 75 kV
Maximum leakage radiation at 1 m from focal spots 0.18 mGy/h
    CHARACTERISTICS Housing C31

  • Lead lined aluminum body.
  • Internal pump for oil circulation, to improve thermal exchange.
  • Three devices are present for thermal safety:
    1 a bimetallic thermal switch, fitted externally on the anode end
    2 a bimetallic thermal switch, internally assembled, series connected with stator common cable
    3 a pressure switch
  • H.T. cable socket: Parker type Compact Taper
  • Solutions for higher dissipation available