X-ray complex CLINOMAT IMX-7

X-ray complex CLINOMAT IMX-7 is a radiographic tilting table with left loading SPOT FILM DEVICE and motorized diaphragm, that operates in 3 positions: Vertical 90°;
Horizontal 0°;
Trandelemburg -15°.

X-ray complex CLINOMAT IMX-7

Table top is very flat and also particularly suitable for traumatized and disable patients .
It can be moved in longitudinal (and transversal but only with the option IMA-71) direction by means of control panel located in the front part of S.F.D.
This equipment allows you to make diagnostic radiographies and also radiographies to standing patients by means of an additional stand (Optional upon request).

Tilting table for general purpose, very strong and built according to the most recent standars.
Silent, smooth and counterbalanced movements with electromagmetic brakes.

Table top with longitudinal run of +50 cm (head-side) and -35 cm (feet-side) in order to make patient positioning end centering easier.
Tilting with progressive start and automatic stop in vertical, horizontal and Trendeleburg positions (-15°).

X-ray complex CLINOMAT IMX-7
X-ray complex CLINOMAT IMX-7
X-ray complex CLINOMAT IMX-7
X-ray complex CLINOMAT IMX-7
X-ray complex CLINOMAT IMX-7
CLINOMAT-1
CLINOMAT-2
CLINOMAT-3
CLINOMAT-4
CLINOMAT-5
X-ray complex CLINOMAT IMX-7 X-ray complex CLINOMAT IMX-7
X-ray complex CLINOMAT IMX-7
X-ray complex CLINOMAT IMX-7
X-ray complex CLINOMAT IMX-7
X-ray complex CLINOMAT IMX-7
X-ray complex CLINOMAT IMX-7
X-ray complex CLINOMAT IMX-7
X-ray complex CLINOMAT IMX-7

Automatic electronic collimator.
Elettronic automatic left loading spot device that can be preset for holding image intensifier of 6″ 9″ 12″ 14″ (optional), and can exclude or can be preset for containing an automatic exposure control (AEC).

It can contain standard cassette from cm 18×24 to cm 35,6×35,6.

It has 24 programs with the following
Divisions: 2, 3, 4, 6.

On the display you can read cassette size, selected division and in-out position.
Rapid exposure might be performed.
Self suspension for image intensifier.
Servomotor for longitudinal movement of spot.

    OPTIONAL:

  • Transversal movement (+/-11 cm) of table -top;
  • 6″ 9″ 12″ 14″ image intensifier with monitor on trolley;
  • Compression band;
  • Shoulder rest;
  • Potter bucky with assembling kit;

See the scheme » Sizes of CLINOMAT IMX-7 (2 ways movements table top) »

See the scheme » Sizes of CLINOMAT IMX-7 + IMA-71 CLINOMAT (4 ways movements table top) »

See the scheme » Sizes of CLINOMAT IMX-7 + IMX-6 (2 ways movements table top + Column with floor rails) »

Specifications X-ray system SIMPLEX IMX-15

DESCRIPTIONDATA
Power supply 230V 50Hz 10A
Weight of the equipmentIMX-7 797kg
TABLE
Table top dimensionsIMX-7
IMX-7+IMA-71
212 x 54 cm
Table top longitudinal travel head-side 50 cm
Table top longitudinal travel feet side 50 cm
Table top lateral travelIMX-7+IMA-71±10 cm
S.F.D. longitudinal travelIMX-7
IMX-7+IMA-71
~88 cm
Bucky travel (OPTIONAL) ~110 cm
Table top height 88 cm
Table top to film distance 11 cm
Focal distancemin 77cm
max 102 cm
Suitable cassette sizes (S.F.D.)from cm 18×24
to cm 35,6×35,6
Absorption coefficient~ < di 0,8 mm Al
Max patient weight 135 kg
S.F.D. Grid36 x 38 cm
Foc. 80 cm – Ratio 10.1 – 150 Pollice
TUBE STAND
Counterbalanced manual vertical movement of tube arm with stop by electromagnetic brake
Tube arm vertical movementMin. 52 cm
Max. 195 cm
Tube arm rotation around horizontal axis +/- 110°
Stand maximum rotation on its vertical axis with mechanical bloc every 90°+/- 180°
Source – film distance max. 113 cm.
COLLIMATOR PREDISPOSITION
Max. Weight9,7 Kg
Check compatibility with X-ray tube 
Lamp’s power supply24V~ with 3 poles cable
(2+ grounding)
ModelsPlease contact us for compatibility check
TUBE PREDISPOSITION 
Max. Weight24 Kg
Check compatibility with collimator 
ModelsPlease contact us for compatibility check
ELECTRIC CLASSIFICATION 
ClassCEI EN 60601-1
(Class I type B) and further corrections
REQUIRED ENVIRONMENTAL FEATURES
Temperature (operative)+10°C to +35°C
Temperature (storage)–10°C to +40°C
Maximum Humidity80% without dew
Maximum Altitude for using3500 meters

This system is completed Dose Area Product Measuring System VacuDAP according to the article 17 the Law of Ukraine «On Protection of People against lonizing Radiation».

X-Ray system can be delivered according to Decree of The Cabinet of Ministers of Ukraine №224 from 20 march 2020.