ANTARIX X-ray Complex IMX-4B

ANTARIX X-ray Complex IMX-4B is a radiographic table with table-top at fixed height with potter bucky, vibrating grid and manual light beam diaphragm.

ANTARIX X-ray Complex IMX-4B

This system enables you to make x-rays to standing patients in case you have a vertical bucky stand.

Device in conformity to 93/42/CEE directive.

See the scheme » Dimensions of ANTARIX X-ray Complex IMX-4B »

Specifications ANTARIX X-ray Complex IMX-4B

DESCRIPTIONDATA
Power supply230V 50Hz 5A
Weight of the equipment465 kg
TABLE
Table top dimensions2200 mm x 746 mm
Table top travel, longitudinal± 407 mm
Table top travel, lateral± 120 mm
Bucky travel515 mm
Motorized Table top height (V=30mm/s)555 / 810 mm
Table top to film distance98 mm
Suitable cassette sizesSee bucky
Absorption coefficient ≤ 0,8 mm Al.
Max patient weight250 Kg
Potter Bucky24V 50Hz 1A
230V 50Hz 1A
Grid480mm x 436mm
103 LPI – R10.1 Foc. 100 cm
TUBE STAND
Counterbalanced manual vertical movement of X-Ray tube with stop by electromagnetic brake.
Transversal movement of X Ray with stop by electromagnetic brake (OPTIONAL)152 mm
Vertical sliding from ground to X-ray centremin. 520 mm
max. 1950 mm
Manual rotation about horizontal axis of tube± 110°
Rotation about vertical axis of tube stand± 90°
Source – film distancemax. 1200 mm
COLLIMATOR PREDISPOSITION
Max. Weight9,7 Kg
Check compatibility with X-ray tube 
Lamp’s power supply24V~ with 3 poles cable
(2+ grounding)
ModelsPlease contact us for compatibility check
TUBE PREDISPOSITION
Max. Weight24 Kg
Check compatibility with collimator 
ModelsPlease contact us for compatibility check
ELECTRIC CLASSIFICATION
ClassCEI EN 60601-1
(Class I type B) and further corrections
REQUIRED ENVIRONMENTAL FEATURES
Temperature (operative)+10°C to +35°C
Temperature (storage)–10°C to +40°C
Maximum Humidity20% to 80% without dew
Maximum Altitude for using3500 meters

ANTARIX X-ray Complex IMX-4B