Rotating anode X-ray tube RTM90 for general radiographic and fluoro procedures, from medium to heavy duty systems.
SPECIFICATIONS X-ray tube RTM90 |
Maximum peak voltage |
150 kV |
Anode angle and diameter |
12.5°-90 mm |
Maximum anode dissipation |
750 W (60000 HU/min) |
Anode heat storage capacity |
225 kJ (300 kHU) |
Rotating anode speed |
3000 – 10000 rpm |
CHARACTERISTICS X-ray tube RTM90
- TZM target with large thermal capacity
- Bearing assembly and bearing lubricant system are designed for long life under hardest x-ray operating conditions
- Focal track in tungsten-rhenium alloy to prevent cracking
- High thermal emittance oxide coating on anode and rotor provides high anode heat dissipation rate
- Severe tests during conditioning assure best performances
- Available also special versions for:
– reloading in GE-CGRand SIEMENS housings (supplied with anode end adaptation piece)
– PHILIPS equipment, to assure full filament emission compatibility
FOCAL SPOTS • mm |
INPUT POWERS • KW |
low speed |
high speed |
0.3/0.6 |
6/24 |
9/35 |
0.6/1.0 |
24/40 |
35/70 |
0.6/1.2 |
24/60 |
35/85 |
0.6/1.3 |
24/62 |
35/90 |
0.6/1.5 |
24/64 |
35/110 |
1.0/2.0 |
40/75 |
70/137 |
|
RELOADING
- CG52 – CGR STATORIX 240
- CG53 – CGR STATORIX 260
- CP52 – PHILIPS ROTALIX 350
- CC52 – COMET D09
|